What TMS RF is
TMS RF is Windows software that turns a USB RF receiver and wireless torque sensors into a station that shows, judges and records every tightening.
The computer receives torque and angle data from the sensors through an RF 2.4G receiver on a USB port. TMS RF shows readings and curves in real time, judges OK or NOK against your target and tolerance, and saves each result in a database on the computer. You can query, export and synchronize results, or pass them to a PLC over Modbus or Open Protocol.
TMS RF only measures, judges and records. It does not control or cut off the power of any machine. An OK result does not mean the tool is calibrated or that the fastener meets the design requirement.
Work modes at a glance
| Mode | Judged by | Typical use |
|---|---|---|
| Torque Wrench | Peak torque | Manual wrench checks |
| Torque & Angle | Angle or torque, by control mode | Tightening with angle control |
| Torque Inspection | Torque when the target angle is reached | Measurement started by angle |
| Power Tool | Peak torque | Pulse, impact and clutch tools |
| Batch Jobs | Steps in order; only OK counts | Several bolts on one workpiece |
Tool Inspection, which calculates capability indices such as Cpk, is covered in the next part.
Data storage
Accounts, settings, sensors and records are stored in the application data folder of the current Windows user, and the installed and portable versions share them. A sensor must be paired in System Settings before a mode can use it. PLC communication, sync and the read API are optional.
Roles and permissions
Each account uses one of three permission templates, and an administrator can adjust single functions per user.
| Template | For | Default abilities |
|---|---|---|
| A (Admin) | Administrators | All functions, including users and sync |
| B (Supervisor) | Engineers, QC, shift leaders | Change work parameters, pair, calibrate; view users only |
| C (User) | Operators | Run work, view and export records; no parameter, pairing or calibration changes |
Installation and first run
TMS RF installs in seconds without administrator rights, and the first start gives you one default administrator account to secure.
System requirements
| Item | Requirement |
|---|---|
| Operating system | Windows 10 or 11, 64-bit |
| Microsoft Edge WebView2 Runtime | Required; it draws the screens |
| Installation rights | A standard user is enough |
| Disk space | About 19 MB for the program; keep 1 GB free for records |
Installer or portable
| File | Size | WebView2 |
|---|---|---|
| Offline installer | 304 MB | Included; no network needed |
| Standard installer | 6 MB | Downloaded during setup if missing |
| Portable program | 19 MB | Must already be installed |
Functions are identical. The portable program installs nothing but uses the same data as the installed version, so never run both at once.


Windows security warning
The files carry no digital signature, so Windows may show an unknown-publisher or SmartScreen warning. Run them only after IT approves, using Run or More info then Run anyway. Do not disable SmartScreen or antivirus. Compare the SHA-256 in SHA256SUMS.txt with the file.
Data location
The main database is %APPDATA%\com.tms.zipptork\static-torque.sqlite3. Each Windows account has its own data, and uninstalling does not delete it.
First run and the default administrator
There is no setup wizard. You get one account, username admin and password admin, template A. Sensor and record lists are empty, Modbus TCP, Modbus RTU and Open Protocol are off, and the read API listens only on 127.0.0.1, port 8080. When you later enable Modbus TCP or Open Protocol, remember they have no encryption or account authentication and belong on a controlled factory LAN only.
The program does not force a password change. Open User Management, click Reset Password on the admin row, and change it at once. It checks only that the password is not empty and matches. There is no recovery, so keep two personal administrator accounts.
Pre-flight checklist
- The login screen appears, and the admin password is changed.
- Each person has a personal account; units and language are correct.
- The receiver and sensors show in Sensor Pairing.
- A trial measurement gives the expected judgement and record.
Login and main screen
You log in with a personal account, then use the left menu and the top bar to move between modes and settings.

Username and password are case-sensitive. After three wrong entries each retry is delayed 0.25 to 2 seconds; the program has not frozen. A login lasts 8 hours, and a permission change forces a new login.
Main screen and navigation

User Management appears only with the View · User List permission. A mode that was never set up opens its settings screen; otherwise it opens the work screen and starts work. Records are opened with Work Records in the top bar.
Leaving a work screen stops work with no confirmation, and a torque application in progress is not saved.
Top bar buttons
| Button | Action |
|---|---|
| Settings (gear) | Opens the mode’s settings and stops work |
| Work Records (clipboard) | Opens records for the current mode |
| Work, Back | On a settings screen, save first, then open the work screen. Back does not discard changes. |
| Toggle light / dark mode | Switches the theme |
Language and formats
There are 16 interface languages, shared by all accounts on the Windows user. Decimal signs follow the language: a comma for Deutsch, Italiano and others, a period for English. A target of 12.5 is typed 12,5 in Deutsch. Dates and times follow the Windows regional format, not the interface language.
First-time setup
Eight steps take an administrator from a fresh install to a first trial measurement and its record.
- Log in as admin and change the password.
- In User Management, use Add User to create accounts with the role User, Supervisor or Admin.
- In System Settings, Global Settings, set the Torque Unit (default N·m) and Language.
- In Sensor Pairing, confirm the receiver shows Connected, put the sensor in pairing mode, choose a channel from 03 to 62, and click Start pairing.
- Only if you correct readings against a reference instrument, add a torque factor (Measured Torque divided by Displayed Torque).
- Open a work mode, select the sensor and complete its settings. Target torque must be 10% to 100% of the sensor spec.
- Make one trial measurement and check OK or NOK.
- Open Work Records and find the trial record at the top.
Common first-time problems
| Situation | Fix |
|---|---|
| Windows blocks the program | Files are unsigned; run them only after IT approval. |
| Receiver not found in Sensor Pairing | Reconnect it, close any program using the COM port, click Rescan. |
| Settings will not save | Check sensor, target torque range and tolerance; read the red message. |
| No permission to save | Template C cannot edit work parameters. Use Supervisor or Admin. |
Torque Wrench mode
Torque Wrench judges each tightening with a manual wrench by its peak torque.
A torque application starts when torque rises above 0 and ends when you release the wrench. TMS RF compares the peak with your target and tolerance and saves a record. The work screen has no Start or Stop button; work starts when you open it and stops when you leave.

Settings
Select the Torque Sensor, then the Torque Unit, then enter Target Torque. Choose Tolerance Type ± or +, and a Tolerance Range from 1 to 30 (%). Station, Part Number, batch counting and a torque multiplier are optional. If Part Number is filled in the settings, records always use it. Click Save Settings to switch to the work screen.
OK/NOK logic
| Tolerance Type | Lower limit | Upper limit |
|---|---|---|
| ± | Target × (1 − tolerance %) | Target × (1 + tolerance %) |
| + | Target | Target × (1 + tolerance %) |
A peak between the limits, inclusive, is OK: green banner, no sound, Part No. cleared. Otherwise NOK: red banner, two warning beeps, Part No. kept for rework.
Operating example
Instruction 25 N·m ±10% gives an OK range of 22.5 to 27.5 N·m. A peak of 24.80 N·m is OK. A peak of 28.10 N·m is NOK; after rework, 25.30 N·m is OK.
Torque & Angle mode
Torque & Angle is for tightening to a seating torque and then turning an angle, judged by angle or by torque depending on the control mode.
Each application records the peak torque and the final angle. The result appears only after saving, so the banner stays Pending until then. Use a sensor that can report angle.

When Seating Torque is above 0, the angle counts from 0 as torque first reaches it. Selecting a sensor fills Seating Torque with 20% of its spec; set it for your part. Target Angle is 1 to 360 degrees.
Control mode decides the judgement
| Control mode | OK when |
|---|---|
| Angle | Peak torque reaches the seating torque and the final angle meets ≥ or ≤ the target angle |
| Torque | Peak torque is in the OK range of the target torque; angle is ignored |
In Angle mode the target torque only draws reference lines.
Operating example
Angle mode, seating torque 20 N·m, target 90 degrees, ≥. A peak of 46.30 N·m at 92.4 degrees is OK. At 84.0 degrees it is NOK. A release at 18.50 N·m never reaches the seating torque, so it is NOK.
Torque Inspection mode
Torque Inspection judges the torque at the moment a joint has turned a set angle past a start torque.
Target Torque (Angle Start) is both where angle counting starts and the centre of the OK range. When the angle reaches Target Angle, the torque at that point is locked as the Angle Torque and judged.

- Tolerance works as in Torque Wrench.
- Release before the target angle: no record, and the banner stays Pending.
- A record is saved only if Angle Torque exceeds half of the torque lower limit.
Example: start torque 20 N·m, ±10%, 15 degrees, so the OK range is 18.0 to 22.0 N·m. An Angle Torque of 21.40 N·m is OK; 23.10 N·m is NOK.
Power Tool mode
Power Tool judges the peak torque of each tightening made by pulse, impact or clutch tools.
The mode does not measure angle. Set Tool Type to match the tool, since it decides when one tightening ends.
| Tool type | For | Tightening ends |
|---|---|---|
| Pulse | Pulse wrenches | After 2 consecutive readings of 0 |
| Impact | Hammer-type impact wrenches | After 2 consecutive readings of 0 |
| Clutch | Tools that shut off at torque | At the first reading of 0 |

Tolerance Type “+ -” accepts target ± tolerance; “+” accepts target up to target × (1 + tolerance %). For pulse and impact tools you can turn on Work time reminder: an amber frame appears 0.5 seconds before Work Time. It is only a visual cue and never changes OK or NOK.
Clutch settings
Click Clutch under Tool Type. Work Time and Work time reminder are hidden, and the title becomes Power Tool-Clutch Work. The result appears as soon as the clutch shuts off.

A hidden reminder keeps its old setting. To avoid an unwanted amber frame, click Pulse, set the reminder to Off, then click Clutch and save. Example: 4.5 N·m ±10% gives 4.05 to 4.95 N·m; 4.62 N·m is OK, 3.98 N·m is NOK.
Batch Jobs
A batch job chains several tightening steps, each with its own mode and sensor, into one procedure the operator completes in order.
Each step has its own mode, sensor, target values, tolerance and Batch Count, the OK results it needs. Example: pre-tighten 5 hub bolts with the wrench, turn 5 by torque and angle, then re-check 5 by torque inspection.
Step types
| Step type | Judged by |
|---|---|
| Torque Wrench | Peak torque within tolerance |
| Torque & Angle | Control mode: angle condition or torque tolerance |
| Power Tool | Peak torque within tolerance |
| Torque Inspection | Angle Torque within tolerance |
Groups and categories
Each batch has a Pset ID (001 to 999), a Batch Number, a Batch Name and an After batch completion action: Restart batch operation or Stop batch operation and return to settings. Every batch belongs to a group; a category is an optional tag. Neither can be deleted on screen.

How an operator runs a job

- Open Batch Jobs. Work starts at the first incomplete step.
- Check Model and S/N against your sensor and the target on the step card.
- Tighten. OK raises the step count. NOK is saved but the step does not advance, so retighten.
- At the Batch Count, the next step and its sensor load automatically.
- After the last step, Batch Complete shows and the batch restarts or returns to settings.
Batch records store the station as WS-01 and the part number as AUTO-BATCH.
Save Job, Work and Back on the settings screen reset all step counts to zero. To view settings without losing progress, leave through the navigation menu and return via Batch Jobs.
Capability Index (Cp, Cpk)
Tool Inspection takes a series of torque samples from one tool against one set of limits and calculates Cp, Cpk, Cm and Cmk, so you can see whether the tool tightens repeatably.
What the test does
Open Tool Capability → Tool Inspection, enter the upper and lower specification limits (USL and LSL), then tighten repeatedly with the tool under test. TMS judges each sample OK or NOK, shows Pass, Marginal or Fail against your acceptance thresholds, and saves every capture as one record.

Three test modes
| Mode | What is measured |
|---|---|
| Torque sensor mode | Peak torque of each tightening, from a paired rotary torque sensor |
| Residual torque mode | Torque from an RF 2.4G measuring device; swing torque, reference torque and angle can be typed in |
| Residual torque + torque sensor mode | Sensor torque is judged; residual torque from another instrument is typed in and saved in the same record |
Statistics and judgement are identical in all three. A new installation uses torque sensor mode; the “Mode” field cannot be changed on screen, so contact your supplier for a residual torque mode.
Capture, indices and judgement
TMS captures the peak as one sample when torque returns to 0; “Confirm capability capture” takes one from the current torque instead. A sample from LSL to USL, limits included, is OK.

- Cpk (shown as CPK): the smaller of (USL − M) ÷ (3 × S) and (M − LSL) ÷ (3 × S) over the latest 20 samples. M is the mean and S the standard deviation.
- Cmk (shown as CMK): the same formula over the latest N samples; N is 30, 50 or 100.
- Cp and Cm, (USL − LSL) ÷ (6 × S), are saved in records only.
An index shows “–” until its window is full. It is Pass at or above its acceptance threshold (default 1.33), Marginal below that if CPK is at least 1.00 or CMK at least 1.33, otherwise Fail.
Rounds
“Start collecting data” ends the round and resets the statistics; saved records stay. A new round also starts when the station, tool, sensor or limits change. “Clear All” only clears the screen and does not restart the evaluation.
Records and Export
Every judged job is saved as a local record that you can query, open, export or delete.
Record types and where to open them
The “Work mode filter” offers “All work modes” plus Torque Wrench, Power Tool, Torque & Angle, Torque Inspection, Batch Jobs and three Tool Inspection types. Open records with “Work Records” (clipboard icon) in a work mode, or from System Settings → Data management → “Records & export”, which shows every mode.

Querying
Filter by “Start Date”, “End Date”, “Station”, “Part Number” and “Result” (All, OK, NOK or Rehit). Dates are compared in UTC but times are shown in local time, so widen the range by a day for a full local day. Each query loads and exports at most the newest 2000 records per type.
“No records match these filters.” means filters hide the data; click “Clear filters”. A red frame with dashes in the statistics bar means the query failed; click “Reload”.
Opening a record
Click a card to see the summary, the saved chart, the judgement conditions, work parameters and sensor information. Raw samples are under “Data Points”, collapsed by default.


Exporting and deleting
- List export (“Export Excel (.xlsx)” or “Export CSV (.csv)”): every record under the current filter, one row each, no samples.
- Detail export (“Excel” or “CSV” in a record): that record with its samples. The Excel file adds a “Sample Data Detail” sheet with a chart.
- The “All work modes” export has no column naming the mode; export each mode separately.
- The red trash can deletes a record and queues the deletion for sync. It cannot be undone on screen and needs “Edit · File Management”.
System Settings
System Settings holds everything that is not a work setting: units and language, RF sensor pairing, calibration factors, the indicator light and the sensor error log.
Each subpage has its own save button, and switching subpage discards unsaved changes.
Global settings
“Torque Unit” (kgf·cm, kgf·m, N·m, ozf·in, lbf·in or lbf·ft; default N·m) sets the unit of records, exports and the read API, not of the work screens. Records are stored in N·m, so changing the unit changes only the display. “Language” offers 16 languages. Both apply at once.
Sensors and RF pairing
The RF 2.4G receiver plugs into USB and appears as a COM port. Every work mode needs a paired sensor.
- Open “Sensor Pairing”. TMS detects the receiver; if not, click “Rescan”.
- Put the sensor in pairing mode and select a channel from 03 to 62, one per sensor.
- Click “Start pairing” and check that the result says the channels match.

“Clear” cannot be undone, and work settings that used that sensor become not set up.
Torque factor and torque multiplier
- Torque Factor = Measured Torque ÷ Displayed Torque. Load the sensor and a reference instrument together, enter the reference reading and click “Save Configuration”. For example, 60.0 ÷ 57.00 gives 1.0526. It applies only to later work.
- Torque Multiplier: enter the Gear Ratio and Torque Ratio of up to three stages. Output torque is sensor torque × total torque ratio; output angle is sensor angle ÷ total gear ratio. Each work setting has its own “Enable torque multiplier” switch.
Indicator light
An external three-color light and buzzer on a USB serial port shows the result. Select the COM port and click “Save Settings”; OK then flashes green and NOK lights red with the buzzer. With no port saved, no light turns on and there is no warning.
Sensor error history
This view-only page lists sensor errors such as torque overload, zero point drift and hardware damage. An error becomes “Cleared” when the sensor stops reporting it; there is no manual clear.

Communication settings
Communication settings let a PLC, MES or tightening management system read TMS results or send commands; TMS always waits for the external device to connect.
Three methods
| Method | Medium and defaults | Typical counterpart |
|---|---|---|
| Modbus TCP | Ethernet; TCP port 1502; Unit ID 1 | PLC, SCADA, PC test tool |
| Modbus RTU | RS-485 or RS-232; COM3, 115200 bps, Slave ID 1 | PLC or HMI with only a serial port |
| Open Protocol | Ethernet; TCP port 4545 | Tightening management system, MES |
All three are disabled after a fresh installation. Modbus TCP and RTU share one register map.
Warning: Modbus TCP and Open Protocol have no encryption and no authentication. Any device on the network can read data or lock the tool. Use them only on a controlled factory LAN.

Selecting “Enable …” and saving starts a service and keeps it starting with TMS; “Start” and “Stop” last only for the current session.
Modbus TCP and RTU
Modbus TCP uses Bind IP Address, TCP Port, Unit ID, Max Clients and Idle Disconnect (default 60 s). TMS creates no Windows Firewall rule, so ask IT to allow the port. To test from another computer, read register 30001: 21581 confirms the link. Modbus RTU needs a USB-to-RS-485 adapter and COM Port, Slave ID, Baud Rate, Data Bits, Parity and Stop Bits that match the PLC.

Open Protocol
TMS acts as a tightening controller. Results are sent only in Batch Jobs mode with a batch selected; each batch is a Pset numbered 001 to 999. The client sends MID 0001 first, then keep-alive MID 9999 within the idle time (default 15 s). It can subscribe to results, select a batch, reset batch progress and lock or unlock the tool.
What a PLC can read or command
Read-only registers (30001 to 30064, FC04) cover live and final torque and angle, target and limits, QC result, batch number and OK count, sensor model, serial number and battery, and status bits. Commands are one FC16 write of 20 registers starting at 40002: apply work parameters, select batch, lock tool, unlock tool or reset current batch. The full register map is in chapter 13 of the manual and can be exported from “Register Map”.
Diagnostics
“Diagnostics & Test” shows Modbus TCP request counters and the latest request and response; it only displays data. A common response is IgnoredUnitId, which means a Unit ID mismatch.

Sync and the read-only API
The Sync settings page has two tabs with opposite data flows: the read-only API lets other systems pull live results, and database sync pushes saved records to a database.
Read-only API
TMS includes an HTTP service that answers GET requests with the live torque, target values, last completed result and QC judgement of the connected sensor. It cannot write or control TMS and has no history. After a fresh installation it is enabled on 127.0.0.1, port 8080, so only programs on the same computer can read it.

- Endpoints:
/api/v1/health,/api/v1/torqueand/api/v1/torque/{sensorId}. - Fields include currentTorque, finalTorque and qcResult (“OK” or “NOK”). Always read “unit”, because it follows the torque unit in Global Settings.
- To let other computers read, set Bind IP Address to a private IPv4 address of this computer. 0.0.0.0 and public addresses are rejected. IT must open the port in Windows Firewall.
Warning: the API uses plain HTTP and needs no password. Keep 127.0.0.1 unless other computers must read it. Changing settings needs “Edit · Server and Sync Settings”, which only Admin has by default.
Database sync
TMS pushes new records and record deletions through an HTTPS gateway to a MySQL or MS SQL database; it never connects to the database directly. Your IT or MES team builds the gateway and provides the https:// URL and an access token.
- Sync is manual: records are sent only when someone clicks “Sync now”, in batches of up to 100.
- If “N records need attention” appears, fix the cause (for example a new token), click “Reschedule”, then “Sync now”.
- The URL and token are encrypted with a key kept in Windows Credential Manager. After a restore on another computer or Windows account, enter them again.
Users and permissions
User Management controls who can run work, change settings, export or delete records and manage other accounts.
The permission model
Every account has a role, and each role sets defaults for 14 function permissions, each Allow or Deny.
| Role | Suited for | Default |
|---|---|---|
| Admin | System administrators | All 14 permissions |
| Supervisor | Supervisors, QC staff, process engineers | All except “Edit · User Management” and “Edit · Server and Sync Settings” |
| User | Operators | Run work, view settings and records, export records, run Tool Inspection; cannot save settings |
You can set one permission of one account differently from its role; changing the role clears these adjustments. The last administrator cannot be demoted or deleted.

Adding a user and setting permissions
- Open User Management at the bottom of the navigation menu and click “+ Add User”.
- Enter a Username, which is also the operator name in records, and a Password. The form has no confirmation field and does not check password strength.
- Choose the Role, then switch single permissions between “Deny” and “Allow” if needed.
- Scroll down and click “Add”.

Saving a role or permission change, or deleting an account, immediately logs that account out and stops its work in progress, so do it at a shift change. A deleted username cannot be reused.
Password reset and security
Only an administrator can change passwords: click the key icon, enter the New Password twice and click “Confirm Change”. A reset does not log out open screens, and a login lasts 8 hours. There is no password recovery.
- Change the default
adminpassword after the first login; TMS does not force it. - Keep at least two administrators.
- Do not delete TMS data files to “reset” accounts: this also erases sensors, settings and records.
Troubleshooting
Note the exact message and the time first, then match it to a problem area below; the manual has a full error message reference.
Finish the current work before changing settings, and check permissions first if a save fails. Do not edit database files, disable SmartScreen or antivirus, or test communication on production equipment.
Common problems
| Area | Symptom | Typical cause and fix |
|---|---|---|
| Installation and startup | SmartScreen or “unknown publisher” warning | The installer has no digital signature. Ask IT to approve it; do not disable SmartScreen. |
| Installation and startup | “Downloading WebView2 Failed”, or the portable program will not open | The WebView2 Runtime is missing. Use the offline installer, or ask IT to install the runtime. |
| Login | Menu items missing, or “You do not have permission to perform this operation.” | The account lacks that permission. An administrator adjusts it in User Management. |
| Sensors and RF receiver | “No RF receiver detected yet” | The receiver is unplugged or its COM port is in use. Replug it, close other programs, click “Rescan”. |
| Work settings and judgement | “Target torque must be between 10% and 100% of the selected sensor capacity” | The target is outside the sensor spec. Correct it or select a sensor with a suitable capacity. |
| Records and export | Export buttons are gray | No export permission, a query is running or has failed, or the dates are reversed. |
| Communication | PLC cannot read; Latest Response shows IgnoredUnitId | The PLC’s Unit ID differs from the one in TMS. Make them equal. |
When you contact support
Fill in the Problem Report Form in Appendix B of the manual, and have these ready:
- The software version (bottom of the login screen) and the install method: offline installer, standard installer or portable.
- The Windows version, date and time, screen name, your steps, the exact message or a screenshot, and the role used. Never send passwords.
- For sensors: the model, serial number and Sensor error history. For communication: the method used and the service status text.
Backup, upgrade and removal
TMS RF has no backup button, so you protect the data by closing the program and copying its data folder. Back up before every upgrade, restore or uninstall, because database updates and data deletion cannot be undone inside the program.
What to protect
| Data | Location | In cold backup |
|---|---|---|
| Accounts, sensors, settings, records, communication and sync settings | %APPDATA%\com.tms.zipptork | Yes |
| Sync encryption key | Windows Credential Manager | No |
| Exported files; remote sync database | Where you saved them; the remote server | No |
Data belongs to the current Windows account. The folder holds all password data and records, so keep backups where only authorized staff can reach them.
Cold backup and restore
- Finish work, log out and close TMS RF completely.
- In File Explorer, copy the whole
com.tms.zipptorkfolder from%APPDATA%to the backup location. - Rename the copy with the date and version, and check that
static-torque.sqlite3has the same size as the original.
To restore, close TMS RF, rename the current folder (for example add “-before-restore”), copy the backup in as com.tms.zipptork and start TMS RF. Restoring replaces all current data, so anything added since the backup is lost. On another computer, enter the sync settings again and recheck COM ports and IP addresses.
Upgrading
An upgrade keeps accounts, sensors, settings and records, but the database is updated automatically at first start and this cannot be undone.
- Installed version: run the new installer and keep the default “Uninstall before installing”. Do not select “Delete the application data” if an uninstall window appears.
- Portable version: replace the old main program with the new one.
- After the first start, check the sensors, settings and records.
- Do not downgrade: older programs ignore per-user Deny settings. To go back, restore the backup made before the upgrade.
Removal
Uninstall from Windows Settings → Apps → Installed apps. The uninstall window has “Delete the application data”, cleared by default, so only the program is removed. For the portable version, delete the .exe; the data stays in %APPDATA%.

Warning: selecting “Delete the application data”, or deleting the data folders by hand, permanently erases all accounts, sensors, settings and records. The next start is treated as a new installation.