STA (RF version) – Residual Torque Verification

STA (RF version) – Residual Torque Verification

 

 

What is required

ZIPPTORK 021 1

 

 

System Architecture

ZIPPTORK 021 2

Procedures

 

 

  1. Insert the Dongle into the NB or Tablet.
  2. Open the TMS software.
  3. Set the Dongle port.
  4. Enter STA mode, then press the “Setup” button to open the “Setup” page.
  5. Select the combination code.
  6. Follow the combination shown in the tool combination code to complete the tool combination (head with STA and the wrench )
  7. Turn on the STA transducer.
  8. Select the STA transducer you want to use (To be the same tool combination setting).
  9. Select the working mode <Torque Verification>.
  10. Set the inspection conditions and accuracy.
  11. Return to the work page.
  12. Set the target torque.
  13. Set the target angle.
  14. Start inspection.
  15. Loosen the STA after tightening to the target angle prompted by the system.
  16. QC judgment.
  17. System complete with work record (operator ID, date & time, target torque, actual torque, inspection conditions & accuracy, and tool combination code).